CAL 2000 LT Calibration System

    Key Features

  • » 
    Hand held, easy to use
  • » 
    Up to 10 hours of gas source use
  • » 
    User defined gas concentration and flow rate outputs
  • » 
    Disposable gas generating cells do not expire and are non-hazardous for transport
  • » 
    Interchangeable calibration gases
  • » 
    Traceable to NIST standards
  • » 
    Rated for use in hazardous areas
The CAL 2000 LT calibration system is a portable, battery powered gas generator specifically designed for calibrating Chlorine, Hydrogen, Hydrogen cyanide and Hydrogen sulphide gas sensors. It is rated for use in hazardous areas.
 
The CAL 2000 LT has a built-in mass flow sensor that measures the flow rate of the instrument. This information is used to control the pump to provide the desired flow rate. The generating gas sources are interchangeable. The microprocessor reads the type of gas from a memory chip embedded in the cell body and then it self-adjusts for each type of gas output. When the unit is powered off, the gas is purged from the unit and the hose, leaving it uncontaminated for the next calibration gas. A small, rotary vane, micro air pump draws in ambient air to blend with the generated gas which permeates out of the cell and onto the gas detector sensor. Four heavy duty C alkaline batteries provide approximately 10 hours of operation and the LCD display is capable of showing the menu structure in English, French, German and Spanish.  
 
To ensure proper operation and to maintain NIST certification, the device should be sent back to the factory on a yearly basis for re-calibration of the flow rate, replacement of the internal charcoal filter and electrolyte replacement.
 

Gas Types

Chlorine (Cl2)
Hydrogen (H2)
Hydrogen sulphide (H2S)
Hydrogen cyanide (HCN)
Source Life: 10 hours or 25 hours
Accuracy: ±10%
Repeatability: ±5%
1 - 25 ppm range available for all gas types, specify actual ppm concentration required at time of order
Choose required flow rate: 0.5 or 1.0 LPM

Sensors

Mechanical

Weight: 1,360 g (3 lb)
Size:
10.8 cm W x  21.59 cm H x  7.62 cm D
(4.25” W x 8.5” H x 3” D)
Tubing (supplied): 1.5 m long x 4.76 mm dia (5 ft long x 3/16” dia) Teflon lined hose
Flow Rate: 0.5 or 1.0 LPM

Electrical

Power Source: 4 alkaline C batteries
Battery Life: approximately 10 hours 
Warm-up Time: approximately 5 minutes

Input / Output

User Interface

Display LCD display
Push Buttons SELECT and POWER push button type switches
Menu Languages English, French, German, Spanish

Environmental

Operating Temperature: 0°C to 50°C (32°F to 122°F)
Humidity: 0 - 100% RH (intermittent use)

Certifications

Conforms to: EC Directive 89/336/EEC in accordance with the provisions of Statutory Instrument 2372
Conforms to: EN 50081-1 and EN 50082-1
Intrinsically Safe: Class 1, Division 1, Groups A, B, C, D
European Standards approval for Zone 0

More Brochures

Product Specifications

Application Brochures

Drawings

Software

Sensor Specification

Engineering Specification Samples

DataSheets

Operation Manual

Part Numbers

Description Code
Battery powered, portable calibration unit comes with a 1.5 m long x 4.76 mm OD (5 ft x 3/16 in OD) Teflon lined hose, nylon carrying case, magnetic screwdriver, batteries, internal charcoal filter and manual. Does not include gas source.  CAL-2000-LT
Chlorine (Cl2) specify ppm and flow rate, 1 year or 10 hours of use  CAL-LT-CL2-10
Hydrogen (H2) specify ppm and flow rate, 1 year or 10 hours of use  CAL-LT-H2-10
Hydrogen sulphide (H2S) specify ppm and flow rate, 1 year or 10 hours of use  CAL-LT-H2S-10
Hydrogen cyanide (HCN) specify ppm and flow rate, 1 year or 10 hours of use  CAL-LT-HCN-10
Chlorine (Cl2) specify ppm and flow rate, 1 year or 25 hours of use  CAL-LT-CL2-25
Hydrogen (H2) specify ppm and flow rate, 1 year or 25 hours of use  CAL-LT-H2-25
Hydrogen sulphide (H2S) specify ppm and flow rate, 1 year or 25 hours of use  CAL-LT-H2S-25
Hydrogen cyanide (HCN) specify ppm and flow rate, 1 year or 25 hours of use  CAL-LT-HCN-25

Bookmark and Share

Contact Us

General Inquiry
Application Support
Quote Request
Maintenance Support
Technical Support
RMA Request
  Warranty
Distribution Opportunity