AST-I Industrial Infrared Transmitter

    Key Features

  • » 
    Dual temperature compensated pyroelectric infrared detectors
  • » 
    Low power consumption
  • » 
    RoHS compliant circuit boards
  • » 
    Fast response
  • » 
    Water / dust tight / corrosion resistant polycarbonate enclosure
  • » 
    Linear analog (4-20 mA) output
  • » 
    Target gases include CO2, Hydrocarbons (including Methane), Nitrous Oxide
  • » 
    Measurement ranges from 0-1000 ppm up to 0-100% volume (sensor & gas dependent)

AST-I is an analog transmitter with high quality, industrial grade, miniature infrared sensors. They are supplied, as a standard in a water / dust tight, corrosion resistant polycarbonate enclosure.

These instruments use high quality, long life, dual beam sensors.The infrared sensor is supplied with a microporous PTFE filter attached to provide splash protection only. This filter is not designed to provide protection to the sensor from application of pressurized water.

APPLICATIONS

  • » 
    Commercial Buildings
  • » 
    Institutional Buildings
  • » 
    Waste Water Treatment Plants
  • » 
    Oil & gas plants
  • » 
    Manufacturing Plants
  • » 
    and many more

Gas Types

Carbon Dioxide (CO2) ICD
Hydrocarbon gases (HC): including Methane (CH4) IHC
Nitrous Oxide (N2O) INI

Sensors

Mechanical

AST-ICD, AST-IHC & AST-INI Industrail Analog Transmitters
Enclosure: Water/dust tight, corrosion resistant polycarbonate
Weight: 255 g (9 oz)
Size: 4.9" x 4.9" x 2.9" (125 mm x 125 mm x 74 mm)

Electrical

Power Requirement: 24 VDC nominal (12 - 30 VDC)
(polarity protected) 5 W maximum
Current Draw: 70 - 100 mA
Outputs: Linear 4 - 20 mA analog
(10 bit resolution with current limit)
Fuse: 500 mA anti-surge in-line fuse for supplied power

Environmental

Operating Temperature: -20°C to 50°C (-4°F to 122°F)
Humidity: 0 - 95% RH non-condensing

Note: Measurements above are sensor dependent.

Certifications


Drawings


Product Specifications

Engineering Specification Samples


DataSheets

Operation Manual

Application Brochure

Quick Application Guide
Quick Application Guide For Fixed Gas Detection SystemsFor Fixed Gas Detection Systems

Miscellaneous Files


AST–I Industrial Infrared Transmitters

Industrial carbon dioxide (CO2), watertight housing, 0-5% volume CO2 AST-ICD-BW
Industrial carbon dioxide (CO2), watertight housing, 0-100% volume CO2 AST-ICD-CW
Industrial carbon dioxide (CO2), watertight housing, 0-100% volume CO2 AST-ICD-DW
Industrial hydrocarbon gases, watertight housing, 0-100% LEL CH4 AST-IHC-BW
Industrial hydrocarbon gases, watertight housing, 0-100% volume CH4 AST-IHC-CW
Industrial nitrious oxide (N2O), watertight housing, 0-1000 ppm N2O AST-INI-EW

Metal Protective Guard

» 16 guage, clear anodized metal protective guard
» Larger size for controllers & self-contained systems
» Smaller size for analog & digital transmitters

Fixed Calibration Kit

» Full selection of calibration gases from stock
» All required calibration accessories for field and shop calibration
» Full service and calibration shop for fixed gas detection instruments


Bookmark and Share

Contact Us

General Inquiry
Application Support
Quote Request
Maintenance Support
Technical Support
RMA Request
  Warranty
Distribution Opportunity